Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18564
Brand Name:Biotronik Pulsar-35 Peripheral Stents, 200mm
Model:Product Codes: 379921, 379941, 379926, 379946, 379931, 379951
Affected:Selected lot numbers
Software version:
Recalling Organisation:Biotronik Australia Pty Ltd (NZ)
44 Cornwall Park Ave
Epsom
Auckland 1051
Contact Information:0800 246 069
Manufacturer:Biotronik AG
Brand Name:Biotronik Pulsar-18 Peripheral Stent, 200mm
Model:Product Codes: 366812, 366817, 366822, 366827, 366832, 366837, 366842, 366847
Affected:Selected lots
Software version:
Recalling Organisation:Biotronik Australia Pty Ltd (NZ)
44 Cornwall Park Ave
Epsom
Auckland 1051
Contact Information:0800 246 069
Manufacturer:Biotronik AG
Issue:The manufacturer has received a higher than expected number of complaints regarding incomplete stent deployments with specific sizes and lots of our Pulsar-18 and Pulsar-35 peripheral self-expanding Nitinol stent systems. Potential health hazards resulting from this type of failure include increased procedure time, vessel wall injury or the need to remove the incompletely deployed stent.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/5/2015