Detail of Recall

Type of Product:Device
Medsafe Reference:18554
Brand Name:Siemens Dimension Vista 500 and Dimension Vista 1500 analysers
Model:SMN: 10488224 & 10444802
Affected:
Software version:3.6.1, 3.6.1_MU3p, and 3.6.1SP1
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics (UK) Ltd
Issue:Siemens has identified 2 issues when using Dimension Vista analysers with specified software versions.
1/ Samples may stop processing without notification on the Dimension Vista 500 or Dimension Vista 1500. This issue is due to a failure to exchange aliquot plates, causing the lane to become unavailable.
2/ Complaints were received on the Dimension Vista 1500 for a series of unexpected results, some of which were not flagged with errors or results that were flagged with errors. It has been determined that under a rare set of conditions, a timing issue may cause a Reagent Server to temporarily lose synchronization during the automatic removal of reagent cartridges from Reagent Server 2 to Waste A Container.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/5/2015