Detail of Recall

Type of Product:Device
Medsafe Reference:18545
Brand Name:ASP EVOTECH Integrated Endoscope Disinfection System
Model:Part Numbers: 50004 & 50004-002
Affected:All systems
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Advanced Sterilization Products Inc
Issue:It has been determined that the EVOTECH® Endoscope Cleaner and Reprocessor (ECR) may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope. The likelihood of this occurrence is low, and ASP has not received complaints or reports of adverse events related to this issue.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/5/2015