| Type of Product: | Device |
| Medsafe Reference: | 18545 |
| Brand Name: | ASP EVOTECH Integrated Endoscope Disinfection System |
| Model: | Part Numbers: 50004 & 50004-002 |
| Affected: | All systems |
| Software version: | |
| Recalling Organisation: | Johnson & Johnson (New Zealand) Limited 507 Mt Wellington Highway Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 922 638 |
| Manufacturer: | Advanced Sterilization Products Inc |
| Issue: | It has been determined that the EVOTECH® Endoscope Cleaner and Reprocessor (ECR) may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope. The likelihood of this occurrence is low, and ASP has not received complaints or reports of adverse events related to this issue. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/5/2015 |