Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18538 |
| Brand Name: | Ossur LP rotate, sizes 25-27 cat.1-6 |
| Model: | Product numbers LRP062, LRF062, LRPU62, and LRFU62 |
| Affected: | Manufacturing Lot Number: HF141022 |
| Software version: | |
| Recalling Organisation: | Christchurch Artificial Limb Centre 330 Burwood Road Christchurch 8083 |
| Contact Information: | 0508 383 383 |
| Manufacturer: | Ossur HF |
| Issue: | Ossur is recalling this device due to the outer tube threads not being machined correctly. Use of the affected product may cause the patient to fall, as the shock module can become loose from the foot blade. The failure could possibly be noticed by users as they will feel looseness on ankle foot device. If use is continued, connection of shock adapter to foot blade might break and user may fall. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Sponsor Control |
| Date Commenced: | 23/7/2015 |