Detail of Recall

Type of Product:Device
Medsafe Reference:18533
Brand Name:Siemens Berichrom Protein C on Sysmex CA-1500, CA-7000, CS-2000i, CS-2100i, and CS-5100 instruments
Model:Catalogue Numbers: OUVV15 & OUVV17
Affected:First affected lot: 45059 (subsequent lots with higher lot numbers might also be affected), plus 44851
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:
Issue:Siemens Healthcare Diagnostics has confirmed that a number of lots can lead to the failure of calibration on Sysmex® CA-1500, CA-7000, CS-2000i, CS-2100i, and CS-5100 instruments.
In case the Berichrom Protein C calibration curve on the above mentioned systems is indicated as being invalid due to the 0% point of the calibration by "reaction curve ERROR", the measurement signal does not meet the instrument's check algorithm for at least a value of 0.001dOD and therefore this point gets flagged.
Follow-up communication
The manufacturer has issued a new protocol disc version with new and enhanced assay applications for Sysmex CS-2000i/CS-2100i Systems. This also includes a new protocol disc version with new and enhanced assay applications for Sysmex CS-2500 Systems and on Sysmex CS-5100 systems.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/7/2018