Detail of Recall

Type of Product:Device
Medsafe Reference:18526
Brand Name:Philips AFFINITI 70 Ultrasound System, without Pediatric Cardiology option
Model:AFFINITI 70
Affected:
Software version:1.0.2 or lower without the Pediatric Cardiology
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Ultrasound LLC
Issue:When AFFINITI 70 Ultrasound System is set up to 'Metric' and weight and/or height is entered, a unit conversion error may result in the incorrect calculation of the patient's Body Surface Area (BSA).
Using patient's weight only to calculate the BSA, which is the usual practice for patients under the age of 2 years, would result in a value that is too high by a factor of about 1.7. A Cardiac Index calculated using that value would be too low by a factor of 1.7. An error of this magnitude may not be readily detected by the user.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/5/2015