Type of Product:Device
Medsafe Reference:18518
Brand Name:Abbott Acetaminophen Reagent
Model:List Number: 2K99-20
Affected:Lots: 4579UQ04, 46121UQ07, 46561UQ08, 46207UQ09, 46953UQ12
Software version:
Recalling Organisation:Abbott Diagnostics Division
Ground Floor, Building D
4 Pacific Rise
Mt Wellington
AUCKLAND 1640
Contact Information:0800 656 233
Manufacturer:Sekisui Diagnostics P.E.I. Inc
Issue:The current Instructions for use reports the Interference from N-acetylcysteine (NAC) evaluated on a commercially available analyser. Using a significance criterion of > 10% variance from control, acceptable results were obtained to a level of 800 mg/L N-acetylcysteine (NAC) in a 104 mcg/mL (688 umol/L) acetaminophen sample; this in vitro analysis was performed approximately two hours after the addition of NAC to a serum pool.
Based on testing performed by Sekisui Diagnostics on an ARCHITECT cSystems instrument, the concentration of NAC at which acceptable acetaminophen results are obtained is 200 mg/L NAC in values from a 109 mcg/mL acetaminophen sample, tested two hours after the addition of NAC to a serum pool.
The package insert is being updated accordingly based on these results.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/5/2015