Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18510 |
| Brand Name: | Philips EPIQ 5 Ultrasound System without pediatic cardiology option |
| Model: | EPIQ 5 |
| Affected: | |
| Software version: | 1.3.2 or lower without Pediatric Cardiology option |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Ultrasound LLC |
| Issue: | When EPIQ 5 Ultrasound System is set up to 'Metric' and weight and/or height is entered, a unit conversion error may result in the incorrect calculation of the patient's Body Surface Area (BSA). When using both height and weight to calculate the BSA, which is the usual practice for patients above the age of 2 years, the result may be too high by a factor of about 10, an error that would be readily identified by the user. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/5/2015 |