Type of Product:Device
Medsafe Reference:18510
Brand Name:Philips EPIQ 5 Ultrasound System without pediatic cardiology option
Model:EPIQ 5
Affected:
Software version:1.3.2 or lower without Pediatric Cardiology option
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Ultrasound LLC
Issue:When EPIQ 5 Ultrasound System is set up to 'Metric' and weight and/or height is entered, a unit conversion error may result in the incorrect calculation of the patient's Body Surface Area (BSA).
When using both height and weight to calculate the BSA, which is the usual practice for patients above the age of 2 years, the result may be too high by a factor of about 10, an error that would be readily identified by the user.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory