Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18490
Brand Name:Siemens RAPIDLab 1260/1265 Systems, D50 Glucose Sensor and D51 Lactate Sensor
Model:10321846, 10491394, 10339910 (RAPIDLab 1260); 10321852, 10470366, 10491395, 10335524 (RAPIDLab 1265)
Affected:All systems
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:If the Glucose or Lactate sensors are not completely seated with the contacts on the instrument, Glucose or Lactate results could be generated when, instead, the instrument should alert the operator with a D50 Glucose sensor error or a D51 Lactate sensor error.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/5/2015
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /