Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18489
Brand Name:Heartware HVAD Pump System
Model:1100, 1101, 1102, 1104, 1205
Affected:All systems
Software version:
Recalling Organisation:Health New Zealand - Te Whatu Ora
9 Ruskin Street
Parnell
Auckland 1052
Contact Information:
Manufacturer:HeartWare International Inc
Issue:Heartware has issued a hazard alert advising physicians and patients of issues that could impact on the performance of the HVAD pump:
1. Bent/damaged power supply port connector pins
2. Failure of the "no power alarm" in the event both external power sources are lost
3. Abnormal power source switching
4. Driveline outer sheath discolouration and cracking over time
5. Snagging or pulling of the driveline
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Public Recall
Date Commenced:15/5/2015
Hide menus