Detail of Recall

Type of Product:Device
Medsafe Reference:18488
Brand Name:Biomerieux Vitek2 AST-YS07 Test Kits
Model:Catalogue numbers: 414967
Affected:Lot numbers: 287314910, 287317310, 287323520, 287336220, 287340140
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:BioMerieux Inc
Issue:An issue related to Out-of-Range-Low (OORL) Flucytosine (FCT) QC results for candida krusei ATCC 6258, the Package Insert QC strain, has been identified in association with the VITEK 2 AST-YS06 and AST-YS07 test kits.
Internal investigation at the manufacturing site has observed the OORL FCT QC results for candida krusei ATCC 6258. Internal testing of FCT with additional isolates shows a trend to call one to two doubling dilutions low. Candida krusei are intrinsically resistant to FCT and are therefore forced "resistant" in VITEK 2 software (as of version 7.01); the VITEK® 2 7.01 FCT forcing rule does not apply to other candida species. User reports have been limited to the ATCC 6258 QC strain; however, impact to patient results for candida krusei and other yeast cannot be ruled out.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/5/2015