Type of Product:Device
Medsafe Reference:18469
Brand Name:Idoman Thermablate Endometrial Ablation System
Model:Product Codes: 22101 & 21004
Affected:All lots
Software version:
Recalling Organisation:New Zealand Medical & Scientific Ltd
Level 4, Building 10
666 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:09 259 4062
Manufacturer:Idoman Teoranta
Issue:The Instructions For Use (IFU) of the Thermablate Endometrial Ablation System (EAS) has been updated following four case reports of uterine perforation. There is an increased risk of patient injury (including damage to non-targeted tissue) if both Thermablate EAS procedure and hysteroscopic tubal occlusion/sterilisation are performed on the same day.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Healthcare Professional
Date Commenced:7/5/2015