Detail of Recall

Type of Product:Device
Medsafe Reference:18454
Brand Name:GE D-Fend & D-Fend+ Water Traps
Model:Part Numbers: 876446-HEL (D-Fend), 881319-HEL (D-Fend+)
Affected:Lots: J9964985, J10201698, J10258248, J10284871, J10284872, J10344108 (D-Fend); J9906966, J9965152, J10258251, J10300071 (D-Fend+)
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Healthcare Finland Oy
Issue:Occlusion of specific lots of D-Fend/D-Fend+ water traps could occur immediately after replacement. This water trap occlusion could cause a loss of respiratory airway gas monitoring. When this issue is detected the connected monitoring device will issue an audio alarm and display "low gas sample flow", "replace water trap" or "sampling line blocked" warning messages. Unattended or prolonged warnings could lead to an adverse clinical situation if the CO2 value would rapidly change in certain clinical conditions.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/4/2015