Detail of Recall

Type of Product:Device
Medsafe Reference:18449
Brand Name:Siemens Immulite 2000 and Immulite 2000 XPi, Reagent Carousel Temperature Sensor
Model:Component: 10290607
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens has identified that affected systems have been manufactured with a defective reagent carousel temperature sensor PCB or may have had a part replacement with a defective PCB assembly.
A particular set of the temperature sensor boards were built incorrectly. The incorrectly assembled temperature sensor boards may cause icing on the reagent carousel casting. If there is icing within the reagent carousel, there is a potential for the reagent carousel to encounter a mechanical jam which will stop the system from processing samples and will generate errors.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/4/2015