| Type of Product: | Device |
| Medsafe Reference: | 18449 |
| Brand Name: | Siemens Immulite 2000 and Immulite 2000 XPi, Reagent Carousel Temperature Sensor |
| Model: | Component: 10290607 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | Siemens has identified that affected systems have been manufactured with a defective reagent carousel temperature sensor PCB or may have had a part replacement with a defective PCB assembly. A particular set of the temperature sensor boards were built incorrectly. The incorrectly assembled temperature sensor boards may cause icing on the reagent carousel casting. If there is icing within the reagent carousel, there is a potential for the reagent carousel to encounter a mechanical jam which will stop the system from processing samples and will generate errors. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/4/2015 |