Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18428 |
| Brand Name: | GE Neuromuscular Transmission Module |
| Model: | E-NMT-01 |
| Affected: | |
| Software version: | Version 1.0 |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Healthcare Finland Oy |
| Issue: | When the E-NMT-01 module is used in conjunction with the ElectroSensor, the Neuromuscular Transmission (NMT) values may indicate a deeper level of muscle relaxation than the actual level of muscle relaxation. In the clinical situation visual movements of the hand are seen after TOF (Train of Four) stimulation, but the patient monitor shows no counts, or counts are not corresponding to the actual amount of movements. This issue may lead to an inadequate dose of muscle relaxants. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/4/2015 |