Detail of Recall

Type of Product:Device
Medsafe Reference:18427
Brand Name:Siemens Ysio Max X-Ray system
Model:
Affected:Version VE10E
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:Images might be lost and corresponding acquisitions have to be repeated under the following circumstances:
1. During an image recovery process on systems with portable detectors (MAX wi-D, MAX mini) the connection of the portable detector is sporadically not re-established.
2. If either the "undo" button on the RAD subtask card or the "previous series" button on the Image subtask card are pressed during image readout.
3. Sporadically, during an automatic or a manual RIS update. This may cause the deletion of the currently used study and all acquired images within this study.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/4/2015