Detail of Recall

Type of Product:Device
Medsafe Reference:18425
Brand Name:Siemens Dimension Vista LOCI Thyroid Stimulating Hormone
Model:Catalogue Number: K6412
Affected:Lot: 14237AA
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens Healthcare Diagnostics has determined that Dimension Vista® LOCI Thyroid Stimulating Hormone (TSH) Lot 14237AA may produce falsely depressed results ranging from -23% to -96% near the end of the well set due to inadequate reagent volume in a Flex® well. Quality Control will detect the issue if it is run near the end of an affected well set.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/4/2015