Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18384
Brand Name:Siemens ADVIA Chemistry Systems, Amylase
Model:Catalogue Numbers: 07498401 & 03031177
Affected:Lot Numbers: 304821 & 304816, Expiry Apr-2015
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens Healthcare Diagnostics has confirmed that Amylase reagent kit lots 304821 and 304816 for use on the ADVIA® 1200, 1650, 1800 and 2400 Chemistry Systems demonstrate an increased incidence of absorbance flags (U or U accompanied by u). Absorbance flags (U) indicate an abnormally high reagent blank absorbance exceeding the reagent blank limit. In some cases with these reagent lots, sample results may also have a Sample Absorbance (u) flag indicating a high absorbance for the sample that occurs in conjunction with the (U) flag. Amylase results can be generated with these flags.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/4/2015