Detail of Recall

Type of Product:Device
Medsafe Reference:18346
Brand Name:Steris Caviwave Pro Ultrasonic Cleaning System
Model:Product Codes: ST-CRP1172, ST-CRP1172401
Affected:All serial numbers
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Steris Corporation
Issue:STERIS has learned through customer feedback and field service experience that should the Caviwave Pro Ultrasonic Cleaning System experience a temporary interruption in electrical power during a processing cycle, it is possible that the circulation pump or ultrasonic generator may stop functioning. The machine does not provide an alert notifying the user of this device malfunction.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/4/2015