Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18335
Brand Name:Uninterruptible Power Supply (UPS) for Philips CT / Advanced Molecular Imaging Systems
Model:BrightView X / XCT, Cardio MD, Forte, Gemini TF/ GXl (6 slice) / TF Astonish / 16 Power / GXL / TF 64 w/ TOF Performance, Precedence, Brilliance CT 16 Slice Air / 64 Channel / Big Bore / iCT / iCT - BETA TEST SYSTEM / iCT SP, Ingenuity CT / Core / Core128
Affected:All units
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:Philips has become aware of UPS device failures that resulted in leaking of electrolyte (sulfuric acid), overheating and/or emission of fumes. Analysis has identified that these failures are most likely due to batteries within the UPS devices being held in service beyond their usable service life - typically three to five years
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/4/2015
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /