Detail of Recall

Type of Product:Device
Medsafe Reference:18322
Brand Name:ASP Cidex OPA Test Strips
Model:
Affected:All lots
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Medos International SARL
Issue:ASP has identified that data for the CIDEX® OPA Solution Test Strips only supports a 10 months shelf life and not the test strip's labeled 12-month shelf life. There is no potential risk to patient safety, as test strips used at the end of shelf life would indicate a false negative (e.g. "Fail" result for solution above the Minimum Effective Concentration).
Recall action type:Recall
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/4/2015