Detail of Recall

Type of Product:Device
Medsafe Reference:18294
Brand Name:Transonic Hemodialysis Monitor
Model:Catalogue Number: HD03
Affected:Lot Numbers: A81225, A91394 & B21812
Software version:
Recalling Organisation:TekMed NZ Pty Ltd
Hilary Square
Orewa
Orewa 0946
Contact Information:0800 456 217
Manufacturer:Transonic Systems Inc
Issue:This recall is being initiated because of customer complaints indicating the inability to obtain measurements due to the device sometimes seeing no flow or irrelevant flow condition when used with patients during dialysis procedures. No reports of injuries have been reported nor are expected due to the nature of this issue.
The problem is readily identified to the user by a zero or a very low flow value displayed on HD03 and/or no measurements produced in other modes (recirculation; access flow; or cardiac output).
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/3/2015