Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18289 |
| Brand Name: | Radiometer AQT90FLEX Immunoassay Analysers |
| Model: | AQT90FLEX |
| Affected: | All serial numbers |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Issue: | RADIOMETER has recently become aware of a potential clinical issue where a measured parameter with a positive result may be reported as being negative. The problem may occur in rare situations where the analyzer uses an incorrect Hct value equal to 0% as a result of sample transport issues. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 30/3/2015 |