Detail of Recall

Type of Product:Device
Medsafe Reference:18289
Brand Name:Radiometer AQT90FLEX Immunoassay Analysers
Model:AQT90FLEX
Affected:All serial numbers
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:RADIOMETER has recently become aware of a potential clinical issue where a measured parameter with a positive result may be reported as being negative. The problem may occur in rare situations where the analyzer uses an incorrect Hct value equal to 0% as a result of sample transport issues.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/3/2015