Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18271 |
| Brand Name: | Siemens Artis Q/ Q.zen systems |
| Model: | All models |
| Affected: | All serial numbers |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Issue: | It is possible that an electrical connection in the equipment cabinet has not been installed correctly. In potential fault scenarios (eg; if several live wires become defective) and under certain conditions, this may prevent a safety mechanism on the system side from taking effect, thus compromising the electrical safety of the system. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/3/2015 |