Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18258 |
| Brand Name: | GE Treadmill |
| Model: | T2100 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems Information Technologies Inc |
| Issue: | A performance issue with the T2100 treadmills can cause uncontrolled walking belt motion during a stress exercise test. This motion can be seen as an unexpected slowing, followed by sudden belt acceleration in either the forward or backward direction. When this occurs, engaging the emergency stop button (ESTOP) will not immediately stop the belt. Rather, it will coast to a stop within 35 seconds when slowing from maximum speed. The sudden change in belt speed and/or direction can result in a fall and injury to the patient. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/3/2015 |