| Type of Product: | Device |
| Medsafe Reference: | 18257 |
| Brand Name: | Medtronic Dual-Chamber External Pulse Generator |
| Model: | 5392 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Issue: | Medtronic has received six (6) reports (out of approximately 6,000 Model 5392 EPGs distributed globally) where the negative terminal of a commercially available AA (LR6) battery did not maintain a sufficient connection with the battery drawer electrical contact. This issue could prevent the EPG from powering on or cause the EPG to abruptly lose primary battery power, potentially stopping delivery of pacing therapy. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/3/2015 |