Detail of Recall

Type of Product:Device
Medsafe Reference:18231
Brand Name:Roche Online Dat Cannabinoids II
Model:Catalogue Number: 04491009190
Affected:Lot numbers: 698113, 604195, 604194, 696131, 698110 & 696133
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Issue:Internal investigations by Roche Diagnostics verified that within the run precision (CV) for the positive/negative control with different reagent lots is approximately 10%. The specification is lower than 8%. Patient sample results which are very close to the cutoff might recover below the cutoff. The affected reagent lots expire in May 2015.
Newer lots are not affected by the precision issue and can be used as a replacement.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/4/2015