Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18211
Brand Name:Getinge 46-Series Washer Disinfector with PACS 300/350
Model:CM302
Affected:All units with PACS 300/350
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Getinge Disinfection AB
Issue:The manufacturer has received two customer complaints from Germany regarding sediment residuals in the manifold of the washer and disinfector of Getinge 46-Series (CM302) with PACS 300/350 control.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/3/2015