Detail of Recall

Type of Product:Device
Medsafe Reference:18197
Brand Name:RaySearch RayStation Radiation therapy treatment planning system
Model:RayStation
Affected:
Software version:3.5, 4.0, 4.5, 4.7
Recalling Organisation:CARSL Consulting
30 Fulford Road
Havelock North 4130
Hastings
Contact Information:09 238 3447
Manufacturer:RaySearch Laboratories AB (publ)
Issue:An issue has been found with the photon dose calculation for Dynamic Multi Leaf Collimation (DMLC) plans for machines where the MLC is positioned above the jaws. The issue applies also to machines with fixed jaws, regardless of MLC/jaw position.
SMLC and VMAT dose calculations are not affected for any MLC configuration.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/3/2015