Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18181 |
| Brand Name: | Biomerieux Nuclisens easyMAG System |
| Model: | Product Code: 200111 |
| Affected: | From EASYMAG02262 to EASYMAG02403, except EASYMAG02401 |
| Software version: | |
| Recalling Organisation: | Medi'Ray New Zealand Ltd 53-55 Paul Matthews Road Albany Auckland 0632 |
| Contact Information: | 0800 633 472 |
| Manufacturer: | bioMerieux Italia S.p.A. |
| Issue: | The manufacturer has received five customers complaints regarding a breakage on the waste trap of the easyMAG instrument. This non conformity allows liquid leakages and air to get into vacuum circuit causing a permanent alarm condition. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/3/2015 |