Detail of Recall

Type of Product:Device
Medsafe Reference:18181
Brand Name:Biomerieux Nuclisens easyMAG System
Model:Product Code: 200111
Affected:From EASYMAG02262 to EASYMAG02403, except EASYMAG02401
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:bioMerieux Italia S.p.A.
Issue:The manufacturer has received five customers complaints regarding a breakage on the waste trap of the easyMAG instrument. This non conformity allows liquid leakages and air to get into vacuum circuit causing a permanent alarm condition.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/3/2015