Detail of Recall

Type of Product:Device
Medsafe Reference:18173
Brand Name:Abbott Architect Anti-HBc II Reagent Kit
Model:List Numbers: 8L44-25, 8L44-30, 8L44-35
Affected:Lot Numbers: 42072LI00, 42072LI01, 42253LI00, 42253LI01, 42255LI00, 42255LI01, 42256LI00
Software version:
Recalling Organisation:Abbott Diagnostics Division
Ground Floor, Building D
4 Pacific Rise
Mt Wellington
AUCKLAND 1640
Contact Information:0800 656 233
Manufacturer:Abbott GmbH & Co. KG
Issue:The ARCHITECT Anti-HBc II assay shows lower calibrator, control and sample RLU values, potentially leading to reduced specificity and increased false reactive results.
Internal testing showed specificity dropped to 99.2% which is below the package insert claim of 99.5% on a blood donor population. Cosnequently there is the potential for false reactive patient results. Note that assay sensitivity is not impacted.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/2/2015