Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18172
Brand Name:Hamilton G5 and S1 Ventilators
Model:Product Code: HL-159001
Affected:
Software version:between V2.40 and V2.41
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:Hamilton Medical AG
Issue:The manufacturer has received reports of the ventilator display feezing. In these cases, the red alarm lamp on top of the ventilator monitor blinks and a high-priority audible alarm is triggered, but no alarm message is displayed. The patient's ventilation continues but the screen is no longer updated and the user can no longer operate the device.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/3/2015