Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18172 |
| Brand Name: | Hamilton G5 and S1 Ventilators |
| Model: | Product Code: HL-159001 |
| Affected: | |
| Software version: | between V2.40 and V2.41 |
| Recalling Organisation: | InterMed Medical Ltd 71 Apollo Drive Albany AUCKLAND 1311 |
| Contact Information: | 0800 333 444 |
| Manufacturer: | Hamilton Medical AG |
| Issue: | The manufacturer has received reports of the ventilator display feezing. In these cases, the red alarm lamp on top of the ventilator monitor blinks and a high-priority audible alarm is triggered, but no alarm message is displayed. The patient's ventilation continues but the screen is no longer updated and the user can no longer operate the device. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 4/3/2015 |