Type of Product:Device
Medsafe Reference:18161
Brand Name:Dameca Anaesthesia Machines
Model:IntelliSave AX700 (866205/10623-00), Dameca MRI508 (866203/10651MRI-00), Siesta iWhispa (866202/10651-00), Siesta iTS (866163/10653-00), Siesta Breasy (866204/10652-00)
Affected:All serial numbers
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Dameca A/S
Issue:If the patient is connected to the anaesthesia machine in a volume controlled ventilation mode, and attempts a large spontaneous breath, it is possible for the patient to create a high negative airway pressure, around the same time as a volume controlled breath (mandatory breath).This pressure will be limited by the mechanical Negative Pressure Limiter (NPL) valve. The NPL valve opens if the patient generates a negative pressure of -5 and -7.5 cmH20 (hPa). This opening pressure of the NPL valve is defined at an inspiratory flow rate of 3.5-4.5 L/min. If the patient creates a higher flow rate than 4.5 L/min, the NPL valve will continue to function but due to the resistance of the flow and time taken to reach the patient, the airway pressure may decrease further.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Sponsor Control
Date Commenced:5/3/2015