Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18148 |
| Brand Name: | BD Q-Syte Luer Access Split Septum Device |
| Model: | Reference: 385100 |
| Affected: | Lot: 4128925 |
| Software version: | |
| Recalling Organisation: | Becton Dickinson Ltd Unit 4 / 110 Mays Road Onehunga Auckland 1061 |
| Contact Information: | 0800 572 468 |
| Manufacturer: | Becton Dickinson Infusion Therapy Systems Inc |
| Issue: | The manufacturer advises that some units of BD Q-Syte Luer Access Split Septum Device have an off-centered slit in the silicon septum resulting in a thinner column wall. When actuated, units with a thinner column wall may tear leading to leakage of insuate/medication or blood from the side vent opening. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 25/2/2015 |