Detail of Recall

Type of Product:Device
Medsafe Reference:18144
Brand Name:Philips HeartStart MRx Monitor/Defibrillator
Model:Models: M3535A, M3536A
Affected:Selected serial number ranges
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Brand Name:Philips HeartStart MRx Monitor/Defibrillator
Model:Models: M3535A, M3536A
Affected:Selected serial number ranges
Software version:
Recalling Organisation:Laerdal (New Zealand) Ltd
79 Apollo Drive
Rosedale
Auckland 0632
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The MRx could reboot once in any operating mode if the following sequence occurs:
1. In Service Mode, the Status Log is cleared
2. Prior to exiting Service Mode, the MRx generates an entry in the Status Log.
3. Service Mode is exited.

The MRx is then in a state that could trigger a reboot at an indeterminate time, potentially causing therapy to be interrupted or delayed. The MRx will restart within 8 seconds.
Note: If a reboot occurs during clinical use, current device settings and patient record are retained.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory