Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18132
Brand Name:S&N Genesis II Non-Porous Tibial Base
Model:71420182
Affected:Bat: 14CT39993
Software version:
Recalling Organisation:Smith & Nephew Surgical Ltd
Unit A 36 Hillside road
Wairau Valley
AUCKLAND 0760
Contact Information:0800 657 799
Manufacturer:Smith & Nephew Inc
Issue:Smith & Nephew advise that a batch of the Genesis II Knee System non-porous tibial base plate was manufactured with a surface finish that did not meet the product specification. This could result in poor cement adherence and the loosening of the base plate. The device may fail, and revision surgery may be required.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Healthcare Professional
Date Commenced:25/2/2015