Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18116 |
| Brand Name: | GE T2100 and T2000 Treadmills, power cord |
| Model: | T2100 & T2000 |
| Affected: | 408930-001 Manf between 1-Jun-2006 and 10-Dec-2014: 408930-002 Manf between 1-Jun-2006 and 19-Jan-2015: 2028038-008 Manf between 1-Dec-2006 and 28-Jan-2015 |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems Information Technologies Inc |
| Issue: | A GE Healthcare internal quality inspection has found that the power cord connecting directly to the T2100 and T2000 Treadmills may not have been assembled according to specifications. If the power cord was improperly assembled and a separate secondary electrical fault condition exists (e.g., a frayed extension or power cord touching the treadmill chassis), this could possibly result in an electrical shock to the patient or operator. There have been no reported incidences of the Treadmill power cord leading to an electrical shock of a patient or operator at this time. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/2/2015 |