Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18099
Brand Name:Philips Allura Xper X-Ray System
Model:Product Codes: 722003, 722006, 720124
Affected:
Software version:R7.2.x
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:Philips Healthcare has discovered through customer complaints and internal testing an intermittent electronic product defect. In certain circumstances, a software error can lead to a situation where the five minute fluoroscopy audible signal does not sound. No injuries attributed to the problem have been reported.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/2/2015