| Type of Product: | Device |
| Medsafe Reference: | 18096 |
| Brand Name: | GE Proteus XR/a X-Ray system |
| Model: | All devices with Wall Stand model number 2260354 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Hualun Medical Systems Co Ltd |
| Issue: | GE Healthcare has recently become aware of a potential safety issue involving the breakage of the steel cable which supports the bucky device in certain wall stands of Proteus XR/a X-ray imaging systems. A fall of a bucky while the system is in use could result in an injury to a patient or operator. There have been no injuries reported as a result of this issue. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/2/2015 |