Detail of Recall

Type of Product:Device
Medsafe Reference:18087
Brand Name:Cook Zenith Alpha™ Thoracic Endovascular Graft
Model:All models
Affected:All lots
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:William Cook Europe ApS
Issue:The instructions for use supplied with the device are being updated in response to recent findings of distal Type I endoleak, migration and aneurysm growth during on-going longer-term follow-up of patients enrolled in the multi-national clinical investigation of the device. To date, no similar events have been reported from markets in which the device is commercially available. These findings from the clinical investigation were found to be associated with thoracic aneurysms (not ulcers) treated with a single proximal component that resulted in a short distal seal length (i.e., less than 20 mm) and likely inadequate (i.e., less than 10%) distal oversizing for the device.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/2/2015