Type of Product:Device
Medsafe Reference:18075
Brand Name:GE Magnetic Resonance Imaging System, magnetic rundown unit (MRU)
Model:All GE MRI systems
Affected:All serial numbers
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems LLC
Issue:The MRU (Magnet Rundown Unit) is a critical safety component of the MR system that quenches the magnet (i.e., makes it non-magnetic) in emergency situations, such as those that can occur when a ferrous (metal) object is brought into the magnetic field.
At certain sites, the MRU may not be connected to the magnet. If the MRU is disconnected, the MRU will not function as expected and will not quench the magnet when the button is pressed. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be connected to the magnet at all times.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/2/2015