Type of Product:Device
Medsafe Reference:18073
Brand Name:Biomerieux Vitek 2 Neisseria-Haemophilus Identification Card
Model:Catalogue Number: 416381
Affected:
Software version:7.01
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:BioMerieux Inc
Issue:A discrepant inoculum density range has been identified for the Neisseria Haemophilus (NH) Identification (ID) Test Card in the VITEK® 2 7.01 Product Information document (Ref: 514740-1EN1). In Section 6 the documented DensiCHEK™ Plus inoculum density range is 1.80 - 2.20 (this is incorrect). The NH Product Information Culture Requirements Table (Table 47) indicates the DensiCHEK™ Plus inoculum density range to be 2.70 - 3.30 (this is correct).
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/2/2015