Detail of Recall

Type of Product:Device
Medsafe Reference:18056
Brand Name:Cordis Empira NC Rx PTCA Catheter
Model:All catalogue numbers 75RxxyyyN
Affected:All unexpired lots
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Creganna-Tactx Medical (USA)
Brand Name:Cordis Empira Rx PTCA Catheter
Model:All catalogue numbers 85Rxx777S
Affected:All unexpired lots
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Creganna-Tactx Medical (USA)
Issue:Through the investigation of the complaints, Cordis and Creganna-Tactx have observed that the balloon of the affected product, which is covered with a hydrophilic coating, may exhibit resistance to the removal of the balloon cover. Weakening of the affected product in the region of the proximal balloon seal may occur when excessive force is applied to remove the balloon cover. Use of a weakened catheter may lead to balloon burst, shaft separation or deflation difficulties, which could result in procedural delay, vessel trauma or surgery. There is no concern for patients who have already been treated successfully with the affected product.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/2/2015