Detail of Recall

Type of Product:Device
Medsafe Reference:18055
Brand Name:Gambro AK200s and AK200 Ultra S Haemodialysis Monitors
Model:
Affected:All units
Software version:
Recalling Organisation:Gambro Pty Ltd (NZ)
5B Target Court
Glenfield
Auckland 0753
Contact Information:021 966 006
Manufacturer:Gambro Lundia AB
Issue:Gambro has received some complaints related to a technical error which will stop all the pumps and close the venous clamp. One section in the Operator's Manuals for the AK 200 S and the AK 200 ULTRA S instructs the user to contact technical service and to restart the blood pump to return the extracorporeal blood to the patient. Restarting the blood pump is however not possible, since the venous clamp is closed and the consequence will be a blood loss for the patient.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/2/2015