| Type of Product: | Device |
| Medsafe Reference: | 18048 |
| Brand Name: | Siemens Advia Centaur Systems Progesterone Kit |
| Model: | Catalogue: 02382928, 01586287, 02177364 |
| Affected: | Lots ending in "268". Expiry date 22-May-2015 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | Siemens Healthcare Diagnostics has confirmed a high bias on results greater than 30 ng/mL (95.4 nmol/L) when compared to the master curve when using affected ADVIA Centaur® Progesterone kit lots. The high bias may cause neat serum samples with results greater than 60 ng/mL (190.8 nmol/L) to recover within the assay range upon dilution (dilution factor applied). On the ADVIA Centaur Systems, for samples greater than 30 ng/mL (95.4 nmol/L), customers may observe an average dilution recovery between 48.3% and 105.0%. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 30/1/2015 |