Detail of Recall

Type of Product:Device
Medsafe Reference:18048
Brand Name:Siemens Advia Centaur Systems Progesterone Kit
Model:Catalogue: 02382928, 01586287, 02177364
Affected:Lots ending in "268". Expiry date 22-May-2015
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens Healthcare Diagnostics has confirmed a high bias on results greater than 30 ng/mL (95.4 nmol/L) when compared to the master curve when using affected ADVIA Centaur® Progesterone kit lots. The high bias may cause neat serum samples with results greater than 60 ng/mL (190.8 nmol/L) to recover within the assay range upon dilution (dilution factor applied). On the ADVIA Centaur Systems, for samples greater than 30 ng/mL (95.4 nmol/L), customers may observe an average dilution recovery between 48.3% and 105.0%.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/1/2015