Type of Product:Device
Medsafe Reference:18027
Brand Name:Philips CDAS Intera MR Systems equiped with Basic Triggering Unit
Model:Part Number 9065 and 9067 Lithium Polymer Fiber Optic Module Batteries (for use with Philips Basic Triggering Set [452213224543], Monitoring Sensor Set [452213224623], Battery Set [452215033673] and Service Parts [453564099151])
Affected:All batteries with date codes 2808 or earlier
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:Affected lithium polymer batteries may fail after being physically dropped from a height (i.e. from waist height to floor). The failed battery may momentarily produce a flame and then emit smoke, which could lead to a thermal burn to patient or operator in close proximity to the battery. The failure of the battery may occur instantly or may be delayed for a short period of time.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/1/2015