Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18021
Brand Name:Carestream Digital Panoramic System
Model:Models: CS 8100 & CS 8100 3D
Affected:All units manufacturered between Jun-2012 and 9-Sep-2014
Software version:
Recalling Organisation:Carestream Health New Zealand Ltd
c/- Quigg Partners
Level 7, 28 Brandon St
Wellington
Contact Information:021 977 215
Manufacturer:Carestream Health Inc (HCIS)
Issue:Carestream Health New Zealand Ltd has received a report that the column of a CS 8100 demonstration device descended unexpectedly as one of the retaining clips that holds the locating pin in place was missing.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Healthcare Professional
Date Commenced:30/1/2015