Detail of Recall

Type of Product:Device
Medsafe Reference:18020
Brand Name:Philips DigitalDiagnost Digital X-Ray
Model:
Affected:
Software version:Release 4.0.0 and 4.0.1
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:The system is designed to emit an acoustic signal (beep) upon termination of an exposure. However, if the system has been powered on for more than 54 hours, the system will no longer emit this signal. If the system does not provide the expected acoustic signal upon the completion of an exposure, the system operator may be unsure that the exposure was correctly obtained, which may result in the decision to repeat an exposure.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/1/2015