Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 18019 |
| Brand Name: | Medtronic InterStim Neurostimulator |
| Model: | Model: 3023 |
| Affected: | All units |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Brand Name: | Medtronic InterStim II Neurostimulator |
| Model: | Model 3058 |
| Affected: | All units |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Issue: | Current labeling indicates the use of cycling improves device longevity under every programming scenario; however, enabling cycling at certain parameter settings may decrease the device battery longevity. Further, anytime cycling is enabled, longevity estimation displayed on the N'Vision® Clinician Programmer (Model 8840) may not be accurate. There is no impact from this issue to therapy delivered or battery level notifications |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 30/1/2015 |