Detail of Recall

Type of Product:Device
Medsafe Reference:18019
Brand Name:Medtronic InterStim Neurostimulator
Model:Model: 3023
Affected:All units
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:Medtronic InterStim II Neurostimulator
Model:Model 3058
Affected:All units
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:Current labeling indicates the use of cycling improves device longevity under every programming scenario; however, enabling cycling at certain parameter settings may decrease the device battery longevity. Further, anytime cycling is enabled, longevity estimation displayed on the N'Vision® Clinician Programmer (Model 8840) may not be accurate. There is no impact from this issue to therapy delivered or battery level notifications
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/1/2015