Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:18013
Brand Name:Maquet Operating Theatre Lights
Model:PowerLED, Hied, Xten, Standop Volista, Hanaulux 3000 range
Affected:Manufactured between March 2012 and July 2014
Software version:
Recalling Organisation:Maquet Australia Pty Ltd (NZ)
12G Andromeda Cres East Tamaki
Auckland 2013
Contact Information:021 874 882
Manufacturer:Maquet SA
Issue:Maquet has received some complaints of grinding on double (2 arms) or triple suspensions (3 arms), with paint possibly chipping from the junction between 2 arms. Paint chips or particles can fall down from the suspension. This may happen during surgical procedures and paint chips may fall inside the surgical field of a patient.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/2/2015